Samantekt:
It is advisable to monitor and regulate inorganic arsenic (iAs) in food and feedstuff. This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry (ICP-MS) for high-throughput screening of iAs content in seafood samples after microwave-assisted extraction with diluted nitric acid and hydrogen peroxide. High concentration of HCl (8 M) for HG along with hydrogen peroxide in samples of a same concentration as used for extraction leads to a selective conversion of iAs to volatile arsine that is released and transported to the detector. A minor contribution from methylarsonate (≈ 20%) was found, while HG from dimethylarsinate, trimethylarsine oxide is substantially suppressed (less than 1% to iAs). Methodology was applied to Certified Reference Materials (CRMs) TORT-3, DORM-3, DORM-4, DOLT-4, DOLT-5, PRON-1, SQID-1 and ERM-CE278k, in some of them iAs has been determined for the first time, and to various seaweed samples from a local store. The results were always compared with a reference method and selectivity of iAs determination was evaluated. An inter-laboratory reproducibility was tested by comparative analyses of six fish and four seaweed samples in three European laboratories, with good agreement of the results. The method of HG-ICP-MS is sensitive (limit of detection 2 μg kg–1 iAs), well suited for screening of large number of samples and selective at iAs concentration levels at which maximum limits are expected to be set into EU legislation for marine samples